PUTRI, SYAHELMA ANGGELINA (2025) ANALISIS PEMANTAPAN MUTU INTERNAL PEMERIKSAAN GLUKOSA DAN KOLESTEROL PADA ALAT MEDICAL SYSTEM MS-380P. Diploma thesis, Universitas Setia Budi.
INTISARI.pdf
Download (5MB)
LEMBAR PERNYATAAN PERSETUJUAN PUBLIKASI.pdf
Download (546kB)
SURAT KETERANGAN CEK PLAGIASI.pdf
Download (160kB)
BAB I.pdf
Download (4MB)
BAB II.pdf
Download (4MB)
BAB III.pdf
Download (4MB)
BAB IV.pdf
Restricted to Repository staff only
Download (4MB)
BAB V.pdf
Restricted to Repository staff only
Download (4MB)
DAFTAR PUSTAKA.pdf
Restricted to Repository staff only
Download (4MB)
LAMPIRAN.pdf
Restricted to Repository staff only
Download (4MB)
Abstract
Pemantapan Mutu Internal (PMI) merupakan bagian penting dari
manajemen mutu laboratorium untuk menjamin akurasi dan presisi hasil
pemeriksaan. Klinik Utama Ultra Medica Boyolali telah melakukan PMI harian
untuk pemeriksaan glukosa dan kolesterol, namun belum dilakukan analisis
mendalam terhadap hasilnya. Penelitian ini bertujuan untuk menganalisis akurasi
dan presisi parameter glukosa dan kolesterol serta mendeteksi adanya kesalahan
sistematik dan acak berdasarkan aturan Westgard menggunakan grafik Levey-
Jennings.
Penelitian ini menggunakan desain deskriptif kuantitatif dengan data
sekunder berupa hasil kontrol kualitas (QC) dari Januari hingga Desember 2024.
Analisis dilakukan terhadap nilai mean, standar deviasi, bias, dan koefisien variasi
(CV), serta mengevaluasi dengan aturan Westgard.
Hasil analisis menunjukkan bahwa pemeriksaan glukosa dan kolesterol di
Klinik Utama Ultra Medica Boyolali pada alat Medical System MS-380P selama
tahun 2024 memenuhi standar kualitas yang baik dengan pelanggaran paling
banyak pada aturan 1₂s yang bersifat warning rule, serta beberapa pelanggaran pada
1₃s dan 4₁s yang perlu perhatian khusus. Kesimpulannya, pemeriksaan glukosa dan
kolesterol dengan alat Medical System MS-380P memiliki akurasi dan presisi yang
cukup baik.
Internal Quality Control (IQC) is a crucial component of laboratory quality
management to ensure the accuracy and precision of test results. Although the
Clinical Laboratory of Ultra Medica Boyolali has routinely performed daily IQC
for glucose and cholesterol examinations, an in-depth analysis of the data has not
yet been conducted. This study aims to analyze the accuracy and precision of
glucose and cholesterol parameters and to identify any systematic or random errors
using Westgard rules and Levey-Jennings charts.
This research uses a quantitative descriptive design with secondary data in
the form of quality control (QC) results from January to December 2024. The
analysis was carried out on mean values, standard deviation, bias, and coefficient
of variation (CV), along with evaluation based on Westgard rules.
The analysis results showed that glucose and cholesterol examinations at
Klinik Utama Ultra Medica Boyolali using the Medical System MS-380P in 2024,
were able to meet good quality standards, with the most frequent violations
occurring in the 1₂s rule as a warning rule, as well as several violations in the 1₃s
and 4₁s rules that require special attention. In conclusion, glucose and cholesterol
examinations using the Medical System MS-380P demonstrated fairly good
accuracy and precision.
| Item Type: | Thesis (Diploma) |
|---|---|
| Uncontrolled Keywords: | Pemantapan Mutu Internal, Glukosa dan Kolesterol, Medical System MS-380P. Internal Quality Control, Glucose and Cholesterol, Medical System MS-380P. |
| Subjects: | Q Science > Q Science (General) R Medicine > R Medicine (General) R Medicine > RA Public aspects of medicine |
| Divisions: | Universitas Setia Budi > Fakultas Ilmu Kesehatan > D4 Analis Kesehatan |
| Depositing User: | Unnamed user with email baa.si@setiabudi.ac.id |
| Date Deposited: | 30 Dec 2025 01:43 |
| Last Modified: | 30 Dec 2025 01:43 |
| URI: | https://eprints.setiabudi.ac.id/id/eprint/173 |